Trilogy 100 Ventilator

Trilogy 100

Ventilator

Find similar products

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

Specifications

Environmental
Environmental
Temperature
  • 5 – 40 °C
Relative humidity
  • 15 - 95 %
Atmospheric pressure
  • 110 to 60 kPa
Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Controls
Controls
Circuit types
  • Active with PAP; passive
Ventilation types
  • Volume control; pressure control; bilevel
Volume modes
  • AC; SIMV (w/PS); CV
Pressure modes
  • CPAP; S; S/T; T; PC-SIMV (w/PS)
IPAP
  • 4 to 50 cm H₂O
Pressure support
  • 4 to 50 cm H₂O
EPAP for active circuits
  • 0 to 25 cm H₂O
EPAP for passive circuits
  • 4 to 50 cm H₂O
CPAP for passive circuits
  • 4 to 20 cm H₂O
PEEP for active circuits
  • 0 to 25 cm H₂O
PEEP for passive circuits
  • 4 to 25 cm H₂O
Pressure support differential
  • 0 to 30 cm H₂O
Physical
Physical
Weight
  • 5 kg (11 lb)
Size
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Alarms
Alarms
Apnea rate
  • Off; 10 to 60 sec
Circuit disconnect
  • Off; 10 to 60 sec
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 50 to 2000 ml
High minute ventilation
  • Off; 0.1 to 99 l/min
Low minute ventilations
  • Off; 0.1 to 99 l/min
High respiratory rate
  • Off; 4 to 80 beats per minute
Low respiratory rate
  • Off; 4 to 80 beats per minute
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 beats per minute
Minute ventilation
  • 0 to 99 l/min
Leak rate
  • 0 to 2000 l/min
Leak rate
  • 0 to 2000 ml
Peak inspiratory flow
  • 0 to 2000 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
Percentage patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Controls
Controls
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 1 to 60 for AC mode; 1 to 60 for all other modes
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Ramp start pressure
  • 0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Ramp length
  • Off; 5 to 45 min
Flex
  • Off; 1 to 3
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 beats per minute