Hospital Respiratory - Ventilation

Respironics V60

Non-invasive ventilator

Because every breath matters and each one is different, patients need a ventilator that follows their breathing pattern, whatever their acuity. The Philips Respironics V60 uses auto-adaptive technology to provide patient synchrony and therapy acceptance. Improved signal processing technology is finely tuned for both adult and pediatric patients.

Contact information

* This field is mandatory
*

Contact details

*
*
*

Company details

*
*
*
*
*

Business details

By specifying your reason for contact we will be able to provide you with a better service.
*
*
Features
Auto-Trak || Marketing leading hospital NIV
Advanced NIV with our patented Auto-Trak technology

Advanced NIV with our patented Auto-Trak technology

Auto-Trak technology is designed to address the specific challenges of noninvasive ventilation. By providing auto-adaptive leak compensation, inspiratory triggering, and expiratory cycling, Auto-Trak delivers optimal synchrony in the face of dynamic leak and changing patient demand. Hospital modes and options include: AVAPS, PCV, CPAP with C-Flex, PPV and Auto-Trak Plus.
Automatic Mask Calibration || Efficient workflow
Save time with automatic mask calibration

Save time with automatic mask calibration

We support your NIV efforts by increasing the speed and ease of treatment initiation. V60’s pre-defined settings for Philips Respironics hospital masks help you save time. You select the setting, and the V60 automatically calibrates flow characteristics for better monitoring and therapy.
Flexible || Excellent therapy
Flexible, easy, upgradeable

Flexible, easy, upgradeable

Designed to include pediatric use and equipped with several modes, the V60 allows you to meet the specific needs of your patients. The high-resolution touchscreen makes operation easy and facilitates waveform interpretation. The internal 6-hour battery enables intra-hospital transport. And, Respi-Link* allows you to upgrade via the internet. *Not available in all markets
Auto-Trak || Marketing leading hospital NIV

Advanced NIV with Auto-Trak

Provides auto-adaptive leak compensation, inspiratory triggering, and expiratory cycling to deliver optimal synchrony in the face of constantly changing leaks and patient demand-alleviating constant bedside supervision and manual adjustments.
Pre-defined mask settings || Efficient workflow

Pre-defined mask settings save time

The Philips Respironics V60 supports the success of NIV by increasing the speed and ease of treatment initiation. Its pre-defined settings for Philips Respironics hospital masks help you save time. You select the setting, and the V60 automatically calibrates flow characteristics for better monitoring and therapy.
CPAP with C-Flex || Excellent therapy

CPAP with C-Flex for greater patient comfort

CPAP (Continuous Positive Airway Pressure) with C-flex offers three levels of flow-based expiratory pressure relief. This option leads to improved sleep quality and patient comfort, adding greater flexibility and improved treatment acceptance.
Multiple treatment modes || Marketing leading hospital NIV

Multiple treatment modes for a wide range of patients

The Philips Respironics V60 helps you meet the needs of a broader patient population. Designed to include pediatric use and equipped with several modes, the V60 allows you to meet the specific needs of your patients.
Expansive color touchscreen || Efficient workflow

Expansive color touchscreen for easy operation

The Philips Respironics V60 has a high-resolution color touchscreen. The high level of visual detail makes operation easy and facilitates waveform interpretation.
PCV || Excellent therapy

PCV increases control

PCV can be used when greater control of the patient's breathing pattern is required.
AVAPS || Marketing leading hospital NIV

AVAPS

AVAPS (Average Volume Assured Pressure Support) maintains a target tidal volume in a pressure limited mode. It provides extra assurance similar to a volume limited mode with the safety of a pressure limited mode.
Standby mode || Efficient workflow

Standby mode supports patient/clinician interaction

Focus on your patients without being interrupted by nuisance alarms by switching the V60 to standby mode.
PPV || Excellent therapy

PPV for variable breathing patterns

Proportional Pressure Ventilation (PPV) provides inspiratory flow and pressure in proportion to the patient's spontaneous effort*, improving patient control over their ventilation**, and improving comfort.
Internal 6-hour battery || Efficient workflow

Transport flexibility

Internal 6-hour battery enables intra-hospital transport
Auto-Trak Plus || Excellent therapy

Auto-Trak Plus for custom treatment criteria

Auto-Trak Plus, an option available in the S/T, PC, AVAPS, and PPV modes, is for the subset of patients who may benefit from a customized titration of triggering and cycling criteria. In certain patients with extremely weak efforts or in others with low compliance, Auto-Trak Plus lets clinicians make finely-tuned adjustments to the base Auto-Trak algorithm to achieve patient ventilator synchrony.
Data communication capabilities || Efficient workflow

Enhanced communications

Data communication capabilities support hospital connectivity
Respi-Link || Efficient workflow

Respi-Link* for efficient upgrades

Through this remote connection tool, you can perform efficient system diagnostics and easily apply upgrades via internet.

Specifications

General
General
Oxygen inlet pressure range
  • 276 to 600 kPa (40 to 87 psig)
Weight with optional battery
  • 10.9 kg (24 lb)
Weight without optional battery
  • 10.0 kg (22 lb)
Dimensions
  • 33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D)
Patient types
Patient types
Suitability
  • Pediatric (≥ 20 kg) and Adult
Electrical
Electrical
AC voltage
  • 100 to 240 VAC
AC frequency
  • 50 to 60 Hz
AC power
  • 300 V·A
Battery operating time
  • 6 hours in normal conditions
Settings
Settings
IPAP
  • 4 to 40 cm H₂O
C-Flex
  • Off, 1 to 3
CPAP
  • 4 to 25 cm H₂O
EPAP
  • 4 to 25 cm H₂O
Max P (PPV maximum pressure limit)
  • 5 to 40 cm H₂O
IPAP
  • 4 to 40 cm H₂O
Max P (AVAPS maximum IPAP)
  • 6 to 40 cm H₂O
Inspiratory time
  • 0.30 to 3.00 sec
Max P (AVAPS maximum IPAP)
  • 6 to 40 cm H₂O
Min P (AVAPS minimum IPAP)
  • 5 to 30 cm H₂O
Min P (AVAPS minimum IPAP)
  • 5 to 30 cm H₂O
Oxygen percentage
  • 21 to 100 %
Ramp time
  • Off, 5 to 45 min
Respiratory rate
  • 4 to 60 beats per minute
Rise time
  • 1 to 5
Triggering and cycling
  • Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6
AVAPS target tidal volume
  • 200 to 2000 ml BTPS
Max E
  • 0 to 100 cm H2O/l
Max E
  • 0 to 100 cm H₂O/l
Max R
  • 0 to 50 cm H2O/l/s
Max R
  • 0 to 50 cm H₂Ol/s
PPV%
  • 0 to 100 %
Max P (PPV maximum pressure limit)
  • 5 to 40 cm H₂O
Max V (PPV maximum volume limit)
  • 200 to 3500 ml
Patient Data
Patient Data
Breath phase/trigger indicator
  • Spontaneous, Timed, Exhale
PIP
  • 0 to 50 cm H₂O
PIP
  • 0 to 50 cm H₂O
Patient/total leak
  • 0 to 200 l/min BTPS
Patient Trigger
  • 0 to 100 %
Respiratory Rate
  • 0 to 90 beats per minute
Ti/Ttot
  • 0 to 91 %
Minute volume
  • 0 to 99.0 l/min BTPS
Tidal volume
  • 0 to 3000 ml BTPS
Waveform window
Waveform window
Pressure waveform
  • 0 - 50 cm H₂O
Flow waveform
  • 240-2401 /min BTPS
Volume waveform
  • 0 - 3,500 ml BTPS
Alarm adjustable range
Alarm adjustable range
Lo Rate (low respiratory rate alarm)
  • 1 - 89 beats per minute
Hi Rate (hi respiratory rate alarm)
  • 5 - 90 beats per minute
Hi Vr (high tidal volume alarm)
  • 200 - 3,500 ml
Lo Vr (low tidal volume alarm)
  • OFF - 1,500 ml
HIP (high inspiratory pressure alarm)
  • 5 - 50 cm H₂O
LIP (low inspiratory pressure alarm)
  • OFF, 1 - 40 cm H₂O
Lo VE (low minute ventilation alarm)
  • OFF, 0.1 - 991 /min
LIP T (low inspiratory pressure delay time)
  • 5 - 60 sec