Ventilation

Trilogy100

Ventilator

Light, versatile, and easy to use, Trilogy100 offers clinicians and patient one of life's greatest qualities - simplicity. The Trilogy100 portable volume and pressure support ventilator is ideal for use at home or in alternative care sites.

Contact information

* This field is mandatory
*

Contact details

*
*
*

Company details

*
*
*
*
*

Business details

By specifying your reason for contact we will be able to provide you with a better service.
*
*
Features
Passive circuit option
Passive circuit option for passive exhalation

Passive circuit option for passive exhalation

The Passive Circuit Option uses single limb circuit with a passive exhalation port or mask with integrated leak. It is compatible with AutoTrak or flow triggering options. This feature provides an estimate of the exhaled tidal volume, as well as leak compensation in both volume and pressure modes.
Active with PAP circuit option
Active with PAP circuit option supports proximal pressure sensing

Active with PAP circuit option supports proximal pressure sensing

The Active with PAP circuit option uses a single limb circuit and a proximal pressure sensor and active exhalation device. The Flow Trigger initiates when the patient’s inspiratory effort creates a flow equal to or greater than the flow sensitivity setting. It provides for an indication of the delivered tidal volume (Vti).
Two circuit options
Two circuit options for simple, flexible use

Two circuit options for simple, flexible use

Interchangeable active and passive exhalation porting blocks give clinicians the flexibility to choose the best available circuit and interface for patients. Trilogy100's internal PEEP, combined with a single-limb circuit, creates a simple application.
Direct View Ventilation Management Reporting Software
Direct View Ventilation Management Reporting Software for prescription management

Direct View Ventilation Management Reporting Software for prescription management

DirectView Ventilation Management Reporting Software streamlines reporting to help identify adjustments necessary to maintain effective treatment and improve patient care. Data can be exported into most statistical applications in an open format for clinicians to share for further analysis. This allows prescription management through the use of a commonly available SD memory card.

Specifications

Environmental
Environmental
Temperature
  • 5 – 40 °C
Relative humidity
  • 15 - 95 %
Atmospheric pressure
  • 110 to 60 kPa
Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Controls
Controls
Circuit types
  • Active with PAP; passive
Ventilation types
  • Volume control; pressure control; bilevel
Volume modes
  • AC; SIMV (w/PS); CV
Pressure modes
  • CPAP; S; S/T; T; PC-SIMV (w/PS)
Pressure support
  • 4 to 50 cm H₂O
IPAP
  • 4 to 50 cm H₂O
EPAP for active circuits
  • 0 to 25 cm H₂O
EPAP for passive circuits
  • 4 to 50 cm H₂O
CPAP for passive circuits
  • 4 to 20 cm H₂O
PEEP for active circuits
  • 0 to 25 cm H₂O
PEEP for passive circuits
  • 4 to 25 cm H₂O
Pressure support differential
  • 0 to 30 cm H₂O
Physical
Physical
Weight
  • 5 kg (11 lb)
Size
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Alarms
Alarms
Apnea rate
  • Off; 10 to 60 sec
Circuit disconnect
  • Off; 10 to 60 sec
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 50 to 2000 ml
High minute ventilation
  • Off; 0.1 to 99 l/min
Low minute ventilations
  • Off; 0.1 to 99 l/min
High respiratory rate
  • Off; 4 to 80 beats per minute
Low respiratory rate
  • Off; 4 to 80 beats per minute
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 beats per minute
Minute ventilation
  • 0 to 99 l/min
Leak rate
  • 0 to 2000 ml
Leak rate
  • 0 to 2000 l/min
Peak inspiratory flow
  • 0 to 2000 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
Percentage patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Controls
Controls
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 1 to 60 for AC mode; 1 to 60 for all other modes
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Ramp start pressure
  • 0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Ramp length
  • Off; 5 to 45 min
Flex
  • Off; 1 to 3
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 beats per minute