Trilogy 200 Portable ventilator

Trilogy 200

Portable ventilator

Find similar products

Versatile, easy-to-use Trilogy 200 provides invasive and non-invasive ventilator support for adult and pediatric patients in the home, hospital and acute-care sites. The pressure- and volume-mode ventilator has proprietary therapy algorithms, connectivity and enhanced triggering sensitivity which decreases the work of breathing. Trilogy 200 helps care teams provide quality respiratory care while meeting patients’ evolving needs.

Specifications

Environmental
Environmental
Temperature
  • 5 – 40 °C
Relative humidity
  • 15 - 95 %
Atmospheric pressure
  • 110 to 60 kPa
Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.5" H)
Weight
  • 5 kg (11 lb)
Controls
Controls
Circuit types
  • Active with PAP; passive; active with flow
Ventilation types
  • Volume control; pressure control; bilevel
Volume modes
  • AC; SIMV (w/PS); CV
Pressure modes
  • CPAP; S; S/T; T; PC-SIMV (w/PS)
IPAP
  • 4 to 50 cm H₂O
EPAP for active circuits
  • 0 to 25 cm H₂O
EPAP for passive circuits
  • 4 to 50 cm H₂O
CPAP for passive circuits
  • 4 to 20 cm H₂O
PEEP for active circuits
  • 0 to 25 cm H₂O
PEEP for passive circuits
  • 4 to 25 cm H₂O
Pressure support differential
  • 0 to 30 cm H₂O
Physical
Physical
Weight
  • 5 kg (11 lb)
Size
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.5" H)
Alarms
Alarms
Apnea rate
  • Off; 10 to 60 sec
Circuit disconnect
  • Off; 5 to 60 sec
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 40 to 2000 ml
High minute ventilation
  • Off; 0.1 to 99 l/min
Low minute ventilations
  • Off; 0.1 to 99 l/min
High respiratory rate
  • Off; 4 to 80 beats per minute
Low respiratory rate
  • Off; 4 to 80 beats per minute
Control accuracy
Control accuracy
Flow trigger sensitivity
  • 1 to 9 l/min
Breath rate
  • 1 to 60 for AC mode; 1 to 60 for all other modes
Tidal volume
  • 50 to 2000 ml
Ramp start pressure
  • 0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Flex
  • Off; 1 to 3
Flow cycle
  • 10 to 90 %
Ramp length
  • Off; 5 to 45 min
Apnea rate
  • 4 to 60 beats per minute
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 beats per minute
Minute ventilation
  • 0 to 99 l/min
Leak rate
  • 0 to 2000 l/min
Leak rate
  • 0 to 2000 ml
Peak inspiratory flow
  • 0 to 2000 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
Percentage patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Controls
Controls
Tidal volume
  • 0 to 2000 ml
Breath rate
  • 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Ramp start pressure
  • 0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
  • Off; 5 to 45 min
Flex
  • Off; 1 to 3
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 BPM