SureSigns VM4

Bedside patient monitor

Product number:

863063

The SureSigns VM series patient monitors are ideal for monitoring, recording, and alarming multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments, including intra-facility transport.

Specifications Show all

SureSigns VM4
Dimensions
26 cm W x 22 cm H x 14.5 cm D (10.2" W x 8.6" H x 5.7" D)
Weight
3.0 kg (6.6 lbs) excluding optional recorder and temperature module
Screen type
21.3 cm (8.4") SVGA TFT-AM LCD display
Display resolution
800 active pixels/line, 600 active lines/frame
Screen active area
170.4 mm x 127.8 mm (6.71" x 5.03")
Screen refresh frequency
60 Hz
Pixel size
0.213 mm (inch)
Viewing angle
approx. 60 °
Alarm Audio Range
45 - 85 dB
System Response Time
1 sec
Electrical specifications
Internal battery
Lithium ion, Smart battery 10.9 V to 11.1 V, 7200 mAhr
Battery operating time (new, fully charged)
4 hr
Internal power supply
100 to 240 VAC
Battery charge time
<4 hr
Power consumption
72 W
Frequency
50/60 Hz
Recorder
Recorder type
Thermal
Paper width
58 mm (inch)
User Selectable Speeds
6.25, 12.5, 25, and 50 ml/s
Environmental specifications
Mechanical Shock
Complies with mechanical shock requirement in ISO 9919 standards for use within the healthcare facility.
Mechanical Vibration
Complies with mechanical vibration requirement in ISO 9919 standards for use within the healthcare facility.
Operating temperature
10 to 40 °C (50 to 104 °F)
Storage temperature (device)
-20 to 50°C (-4 to 122 °F)
Storage temperature (device plus accessories)
-20 to 40°C (-4 to 104 °F)
Operating humidity
Up to 80% RH, non-condensing
Storage humidity
Up to 90% RH, non-condensing
Altitude
Up to 3000 meters above sea level
ECG
Heart rate range
15 to 300 beats per minute
Heart rate accuracy
Greater of ±10% or ±5 bpm
Bandwidth (normal monitoring)
0.67 to 40 Hz
Bandwidth (filtered monitoring)
0.67 to 20 Hz
Leads
3-lead
Display sweep speeds
12.5, 25, and 50 ml/s
Pacemaker detection
Indicator on waveform display (user-selectable)
ECG size (sensitivity)
2.0, 1.0, 0.5 or 0.25 cm/mV or Auto
Lead Off condition
Detected and displayed
Input impedance
Signal reduction is less than 20% in the 0.67 to 40 Hz bandwidth
Input signal range
±5 mV
Impedance Respiration
Technique
Trans-thoracic impedance
Measurement range
0 to 150 r/min
Accuracy
±1 rpm in the range of 0 to 120 rpm; ±2 rpm in the range of >120 rpm
ECG leads used
RA to LL
Display sweep speeds
6.25, 12.5, 25, or 50 ml/s
Lead Off condition
Detected and displayed
Non-invasive Blood Pressure
Technique
Oscillometric using stepwise deflation pressure
Adult measurement range - systolic
30 mmHg to 270 mmHg (4.0 kPa to 36.0 kPa)
Adult measurement range - diastolic
10 mmHg to 245 mmHg (1.3 kPa to 32.7 kPa)
Initial cuff inflation - Adult
160 mmHg (21.3 kPa)
Adult measurement range - MAP
20 mmHg to 255 mmHg (2.7 kPa to 34.0 kPa)
Adult measurement range - pulse rate
40 to 300 beats per minute
Pediatric measurement range - systolic
30 mmHg to 180 mmHg (4.0 kPa to 24.0 kPa)
Pediatric measurement range - diastolic
10 mmHg to 150 mmHg (1.3 kPa to 20.0 kPa)
Initial cuff inflation - Pediatric
140 mmHg (18.7 kPa)
Pediatric measurement range - MAP
20 mmHg to 160 mmHg (2.7 kPa to 21.3 kPa)
Pediatric measurement range - pulse rate
40 to 300 beats per minute
Neonatal measurement range - systolic
30 mmHg to 130 mmHg (4.0 kPa to 17.0 kPa)
Neonatal measurement range - diastolic
10 mmHg to 100 mmHg (1.3 kPa to 13.3 kPa)
Neonatal measurement range - MAP
20 mmHg to 120 mmHg (2.7 kPa to 16.0 kPa)
Neonatal measurement range - pulse rate
40 to 300 beats per minute
Blood pressure accuracy
8 mmHg maximum standard deviation; ±5 mmHg maximum mean error
Pulse rate accuracy
40 to 100 bpm: ±5 bpm; 101 to 200 bpm: ±5% of reading; 201 to 300 bpm: ±10% of reading
Initial cuff inflation - neonatal
100 mmHg (13.3 kPa)
NBP intervals
Automatic measurements at intervals of 1, 2, 3, 5, 10, 15, 30, 60, 90, 120 minutes, and STAT
Temperature Measurements
Measurement range - monitored mode
26.7 to 43.3 °C (80 to 110 °F)
Measurement range - predictive mode
34.4 to 40.6 °C (93.9 to 105 °F)
Heating/Cooling transient response time
≤150 sec
Accuracy (monitored mode)
±0.1 °C (±0.2 °F)
Measurement time
4 s (Oral); <16 s (Axillary and rectal) sec
Resolution
0.1 °C (0.2 °F)
SpO2
Measurement range
0 to 100 %
Pulse rate
30 to 300 beats per minute
SpO2 accuracy
70 to 100% (exact accuracy depends on sensor) %
Pulse rate accuracy
Greater of 2% or ±1 bpm
Wavelength range
500 to 1000 nm for all specified sensors nm
Maximum optical output power
≤15 mW for all specified sensors mW