Achieva 3.0T TX

MR System

Product number:

781345

Philips Achieva 3.0T TX MR system features unique MultiTransmit technology.It takes high-field MRI to the next level of clinical excellence. Allowing you to confidently expand into the 3.0T growth areas of breast, body and cardiac imaging.

Specifications Show all

Xtend Magnet System
Open bore diameter
70 cm
Cryogen boil-off rate
0 l/hr
Magnet weight
4600 kg
Maximum FOV
55 cm
Typical homogeneity at 55x55x50 cm
≤ 5 ppm
Typical homogeneity at 50 x 50 x 45 cm
≤ 1.8 ppm
HeliumSave technology
(Zero boil-off) Yes
Xtend Gradient - 3.0T Omega HP
Max. amplitude for each axis
45 mT/m
Max. slew rate for each axis
200 T/m/s
Resolution Parameters - 3.0T Omega HP
Max scan matrix
1024 (2048 optional)
Highest in-plane resolution
5 µm
RF Transmit
Parallel RF transmission
Yes, MultiTransmit
RF control
Patient-adaptive. Fully flexible for each source, real time feedback loop
dStream RF Receive
Number of independent receive channels
Channel Independent
Signal chain from coil to reconstructor
Digital signal chain from coil electronics to connector
Digital signal chain from connector to magnet
Digital signal chain from magnet to reconstructor
Location of analog-to-digital converter (ADC)
Inside the coil
Patient Environment
Patient aperture
70 cm
Flare on both ends
Yes
Tunnel diameter at both ends
95 cm
Maximum weight capacity*
250 kg (550 lbs)
Patient transport system
FlexTrak
Patient transport system (optional)
FlexTrak Mammo
Wireless patient physiological synchronization
Yes
Various acoustic noise reduction solutions
Yes
Site Planning
Total gantry installed weight
≤ 5800 kg
Minimum siting requirement
30 m²
dStream Workflow (FlexStream)
FlexCoverage posterior coil
Yes
FlexCoverage anterior coil (optional)
Yes
FlexConnect connectors
Yes
FlexTrak tabletop
Yes
FlexCaddy coil storage (optional)
Yes
FlexTrak patient transport system (optional)
Yes
FlexTrak Mammo mammography solution (optional)
Yes
SmartAssist Efficiency Assistance
SmartStart
Yes
SmartSelect
Yes
SmartExam (optional)
Yes
SmartLine (package dependent)
Yes
SmartLink (package dependent)
Yes